DRJ Dent Res J Dental Research Journal Dental Research Journal 1735-3327 2008-0255 Medknow Publications Pvt Ltd India DRJ-12-231 Original Article The effects of aging process and preactivation on mechanical properties of nickel-titanium closed coil springs Alavi Shiva Haerian Alireza Dental Materials Research Center and Department of Orthodontics, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, Iran Department of Orthodontics and Dental Students Research Center, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, Iran Address for correspondence:Alireza Haerian, Dental Students Research Center, School of Dentistry, Isfahan University of Medical Sciences, Isfahan 8174673461, Iran haerian.alireza@gmail.com May–Jun 2015 12 3 231 234 Copyright: © Dental Research Journal 2015

This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Background:

The most favorable feature of nickel-titanium (Ni-Ti) alloys in orthodontics is producing constant forces in a wide deformation range. Ni-Ti closed coils produce compressive strength and can be used for several purposes such as space closure. The aim of this study was to evaluate the effects of the aging process (temperature changes and prolonged strain) on Ni-Ti closed coils and to assess the effects of preactivation on forces generated by these coils.

Materials and Methods:

A total of 60 Ni-Ti closed coils (G&H Wire Co., Indiana, USA) were divided into four groups (n = 15). Two groups were kept in room temperature and two were incubated in 37°C in normal saline for 45 days. All the samples were extended to 30% of their original length. One group of the incubated coils and one group of room temperature coils were preactivated the same amount of 30%. The incubated groups also received 1000 thermocycles on days 22 and 45. The unloading forces were measured by a universal testing machine on days 0, 22 and 45. The data were analyzed using SPSS. Significance was set at 0.05.

Results:

The unloading forces of the Ni-Ti closed coils was not affected by prolonged heat and moisture either in preactivated groups (P = 0.8) or the nonpreactivated groups (P = 0.6). Furthermore, preactivation had no significant effect on the unloading forces of Ni-Ti closed coils (P = 0.7).

Conclusion:

Within the limitations of this study, both aging process and preactivation do not affect on mechanical properties of nickel-titanium closed coil springs. Preactivation had no effect on the forces generated by Ni-Ti closed coils.

Closed coil springs mechanical properties nickel-titanium
</sec> <sec> <title>Introduction

Tooth movement requires force application. This force is produced by orthodontic appliances. 1 Continuous forces lead to the most efficient tooth movement. Nickel-titanium (Ni-Ti) alloys can generate relatively continuous forces. 2 After the introduction of Ni-Ti alloys to dentistry in 1970, it has been widely used. 3 Producing constant forces in a wide deformation range is the most favorable feature of Ni-Ti alloys, which is due to the two-phase structure of the alloy; the martensite and the austenite phases. 2

For the best use, the alloy must be in the austenite phase when placed in room temperature. 2 Various orthodontic appliances are produced from Ni-Ti alloys such as closed or open coil springs. 4 Open coils generate expansive forces, while closed coils generate compressive forces. 4 Ni-Ti coil springs, unlike elastomeric chains, do not undergo rapid force degradation and unlike stainless steel coils do not generate extremely high forces. 5 Temperature and strain could lead to phase change in Ni-Ti alloys. 6 Therefore, the temperature changes, which occur in the oral cavity may affect the forces generated by the coils. 6 Several studies have showed that prolonged strain and heat could reduce the predictability of forces produced by coils. 1 , 5 , 7 It has been reported that preactivation can prolong the continuity of the forces produced by elastomeric chains. 8 , 9 , 10 The aim of this study was to evaluate the effects of the aging process (temperature changes and prolonged strain) on Ni-Ti closed coils and also to assess the effects of preactivation on the forces generated by these coils.

Materials and Methods

In this experimental in vitro study, sixty 9 mm medium Ni-Ti closed coils (G&H Wire Co., Indiana, USA) were divided equally in to four groups (n = 15). In the 1 stgroup, the samples were stretched to 12 mm (3 mm activation) using a universal testing machine (walter + bai, Switzerland). After that, the samples were returned to their original length by the testing machine, with the crosshead speed of 0.5 mm/s. The forces were recorded at every 0.5 mm. Holding jigs were designed; it was comprised of stainless steel rods, which were placed at 12 mm distances Figure 1. The samples were carefully placed on the holding jig after every test to maintain their 12 mm length during the study. These measurements were repeated in T1 = day 22 and T2 = day 45. In the 2 ndgroup the process was similar to the 1 stgroup except that after the measurement, the samples that were placed on the holding jig were inserted in normal saline (Samen, Mashhad, Iran) then transferred to an incubator (Behdad, Tehran, Iran) and kept in 37°C. Before repeating the tests in T1 and T2 the 2 ndgroup samples underwent 1000 thermocycling cycles from 5°C to 55°C. The 3 rdand 4 thgroup samples were preactivated by the amount of 30% (3 mm) using the gauge provided by the manufacturer (G&H Wire Co., Indiana, USA). In the 3 rdgroup, the process was similar to the 1 stgroup and after the measurements, the samples were carefully placed on the holding jig and kept in room temperature and then the measurements were repeated in T1 and T2. In the 4 thgroup, after the initial measurements the samples were placed on the holding jig and kept in normal saline inside the incubator. Before repeating the measurements in T1 and T2 the samples underwent 1000 thermo cycling cycles between 5°C and 55°C. Finally, the data was analyzed by the repeated measure two-way ANOVA using the SPSS software version 13 (SPSS, Chicago, IL, USA). The significance was set at 0.05.

The designed holding jig.

Figure 1

Results

Statistical analysis showed that the mean forces produced in all groups were slightly higher than the target force stated by the manufacturer (120 gf). The mean forces are shown in Table 1and Figure 2. Comparing the nonpreactivated groups showed that there was no significant difference between the forces in the dry and incubated groups in different time periods (P = 0.6). In the preactivated groups, there was no significant difference between the forces in the dry and incubated groups in different time periods (P = 0.8). Regarding preactivation, the mean forces of preactivated coils (dry and incubated) had no significant difference with the mean forces produced by nonpreactivated coils (dry and incubated) (P = 0.7). In the incubated groups, the mean forces slightly decreased from T0 to T2, but it was not statistically significant (P = 0.6), but in the dry groups, the mean forces slightly increased, but this also was not statistically significant (P = 0.4).

The mean values of unloading forces in different groups. PI = Prestretched and incubated, PD = Prestretched in dry condition, NPI = Nonprestretched and incubated, NPD = Nonprestretched in dry condition (the mean values of unloading forces are presented in gf.), T0 = Tested on day 1, T1 = Tested on day 22, T2 = Tested on day 45.

Figure 2 {Table 1}

Discussion

In this study, we evaluated the effects of aging and preactivation on the forces produced by Ni-Ti closed coils. We measured the forces in the days 0, 22 and 45, while two groups were kept in room temperature and two groups were incubated in 37°C. We found that the mean forces produced by the coils, in all time periods, were slightly higher than the target force stated by the manufacturer. Vidoni et al., 1 Bezrouk et al. 4 and Maganzini et al. 6 also reported that the mean forces produced by Ni-Ti coils were higher than the target force of the manufacturer. This difference could be due to different method of measurement between these studies and the manufacturer. In this study and the studies mentioned, the force measurements were performed by the highly accurate universal testing machine.

The results of this study showed that the forces produced by coils in room temperature had no significant difference with the forces produced by coils kept in the incubator, which means that prolonged moisture and heat did not affect the forces produced by Ni-Ti closed coils. Vidoni et al. 1 and Wichelhaus et al. 3 showed that the presence of moisture and heat doesn′t affect the forces produced by Ni-Ti coils. On the other hand, Tripolt et al. 7 stated that heat increases the forces produced by Ni-Ti coils, but in the oral environment the magnitude of this force change is not significant. Bourke et al. 8 showed that Ni-Ti closed coils could maintain their force levels during oral environment changes. It seems that in normal conditions, phase change does not occur in Ni-Ti alloys present in the oral cavity. Wichelhaus et al. 3 have reported that some manufacturers preactivate the Ni-Ti coils, which results in a longer plateau in the stress-strain charts of the coils. Our study showed that preactivation doesn′t affect the forces produced by Ni-Ti closed coil in either dry or incubated environments. Regarding the change in forces produced by Ni-Ti closed coils, our study showed that the incubated coil (either preactivated or nonpreactivated) forces decreased over time, but the coils kept in room temperature (preactivated or nonpreactivated) showed increased forces over time. However, none of these changes were statistically significant. Nattrass et al. 11 have reported an increase in forces produced by Ni-Ti coils over time whilst, Angolkar et al. 12 showed a decrease in Ni-Ti coil forces over time. Vidoni et al. 1 reported that nitinol coils had no significant force change overtime (after 45 days) but Ni-Ti coils had a slight force decrease after 45 days which was statistically significant but didn′t have clinical significance.

In the studies of Nattrass et al. 11 and Angolkar et al. 12 thermocycling was not performed, which could explain the different results. In Vidoni et al. 1 study, the coils were stretched 50% of their initial length, whilst in our study, the coils were stretched 30% and a different brand of coil was used in their study, which could be the cause of different results.

Conclusion

Aging process does not affect the forces produced by Ni-Ti closed coil spring.

Preactivation has no statistically significant effect on the forces generated by Ni-Ti closed coils.

Vidoni G Perinetti G Antoniolli F Castaldo A Contardo L Combined aging effects of strain and thermocycling on unload deflection modes of nickel-titanium closed-coil springs: An in-vitro comparative study Am J Orthod Dentofacial Orthop 2010 138 451 7 Thompson SA An overview of nickel-titanium alloys used in dentistry Int Endod J 2000 33 297 310 Wichelhaus A Brauchli L Ball J Mertmann M Mechanical behavior and clinical application of nickel-titanium closed-coil springs under different stress levels and mechanical loading cycles Am J Orthod Dentofacial Orthop 2010 137 671 8 Bezrouk A Balský L Smutný M Nosek T Záhora J Hanus J Thermo-mechanical properties of NiTi closed coil springs - force degradation and force regeneration over time, viscous properties Acta Medica (Hradec Kralove) 2013 56 41 6 Espinar-Escalona E Llamas-Carreras JM Barrera-Mora JM Abalos-Lasbrucci C Gil-Mur FJ Effect of temperature on the orthodontic clinical applications of NiTi closed-coil springs Med Oral Patol Oral Cir Bucal 2013 18 e721 4 Maganzini AL Wong AM Ahmed MK Forces of various nickel titanium closed coil springs Angle Orthod 2010 80 182 7 Tripolt H Burstone CJ Bantleon P Manschiebel W Force characteristics of nickel-titanium tension coil springs Am J Orthod Dentofacial Orthop 1999 115 498 507 Bourke A Daskalogiannakis J Tompson B Watson P Force characteristics of nickel-titanium open-coil springs.Am J Orthod Dentofacial Orthop 2010;138:142 e- 7 Force characteristics of nickel-titanium open-coil springs Am J Orthod Dentofacial Orthop 2010;138:142e1-7 Han S Quick DC Nickel-titanium spring properties in a simulated oral environment Angle Orthod 1993 63 67 72 Schneevoigt R Haase A Eckardt VL Harzer W Bourauel C Laboratory analysis of superelastic NiTi compression springs Med Eng Phys 1999 21 119 25 Nattrass C Ireland AJ Sherriff M The effect of environmental factors on elastomeric chain and nickel titanium coil springs Eur J Orthod 1998 20 169 76 Angolkar PV Arnold JV Nanda RS Duncanson MG Jr Force degradation of closed coil springs: An in vitro evaluation Am J Orthod Dentofacial Orthop 1992 102 127 33