This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
The “All-on-four” concept for treatment of edentulous arches incorporates four implants that are placed in between mental foramina in the mandible. The prosthetic framework is an important parameter in stress/strain concentration at the implants, prosthesis, and the underlying bone. Materials such as titanium, zirconia, and carbon fibers have been used for fabrication of framework in the past. The aim of this study was to analyze the effect of framework materials in the “All-on-four” implant system.
Finite element three-dimensional (3D) model of edentulous mandible was simulated using a computerized tomographic scan data of an edentulous patient. Threaded implants were replicated along with the abutments using 3D modeling software and the framework was designed and simulated using material properties of titanium, zirconia, and polyetheretherketone (PEEK). Axial and nonaxial load of 200 N was applied at the abutment region of right distal implants. The computer-generated numerical values were tabulated and analysed by ANSYS software.
Principal strain, von Mises stress and micromotion were assessed in the peri-implant bone region to evaluate its stress condition. Zirconia framework showed the least stress/strain values at axial and oblique loading. Maximum strain values were seen at the PEEK framework material. Zirconia framework in all models showed the least micromotion/displacement.
The stress distribution pattern at implant–bone interface was influenced by the framework material used. The framework material, loading site, and direction of forces influenced the stresses and displacement at the bone–implant interface.
In edentulous patients, implant-supported full-arch restoration is a common treatment option. An implant system called “All-on-four” used for treating such patients incorporates four implants that are placed between the mental foramina region of the mandible. In this system, two implants are placed axially in the anterior region, while two are placed in the posterior region at an angle. All the four implants are connected through a superstructure framework.
Materials used for the prosthetic superstructure framework play a significant role in its biomechanical success. A rigid framework allows absorption and balanced distribution of stresses to prevent deformations.
Recently, a new material called polyetheretherketone (PEEK) has been introduced which exhibits excellent mechanical properties and is used in the fabrication of removable dentures, fixed restorations, dental implants, and implant abutments.
Studies have suggested that stiffer materials show higher stress values in the prosthetic framework as compared to less rigid materials.
The three-dimensional (3D) finite element method (FEM) is a numerical procedure analyzing structures using computer models of a material. It uses a complex system of points (nodes) and elements which makes a grid called a mesh. The mesh is programmed to contain the material and structural properties (Young's modulus and Poisson's ratio), and the design is stressed and analyzed for specific results.
The purpose of the present study was to analyze the effects of different materials used for framework fabrication on “All-on-four” implant system using FEM. The three framework materials used in the study varied in their material properties and were critical for the evaluation of stresses in peri-implant region. Evaluation of various stress distribution patterns for axial and nonaxial stresses and micromotion that occurs around these implants was also assessed.
Designing a finite element model of “All-on-four implant” system
A 3D model of edentulous mandible was simulated using a computerized tomographic scan data of an edentulous patient. Threaded implants (11.5 mm, 13 mm length, and 4 mm diameter) were replicated along with the abutments using 3D modeling software. The bone segments were modeled as cortical bone representing the outer shell of 2-mm thickness and the inner volume of cancellous bone. The implants had 100% bone–implant contact. Following the “All-on-four” concept, two anterior implants were placed at the lower lateral incisor region, while the posterior implants were angled distally (30°) and placed anterior to the mental foramen
Mandible model with four implants according to “All-on-Four” concept.
Designing of a framework using material properties of titanium, zirconia, and polyetheretherketone
The framework was designed as a solid bar of height 5 mm and width 6 mm following the shape of the mandible. The cantilever extension was 11.5 mm. At the end of the bar, 2-mm diameter circle was created on the outer occlusal surface to standardize the load application. The framework was simulated using material properties of titanium, zirconia, and PEEK, obtained from previous studies. Young's modulus and Poisson's ratio were assigned to each of the components
Standardized load application on the outer occlusal surface of framework.
Solid elements were used to represent the geometry. All the 3D models of the implants and framework were built using Solid Edge software (Siemens PLM Software, USA). Bone was built by reverse engineering processor. Meshwork was prepared using HyperMesh software. The analysis was done by ANSYS software (USA).
Application of load on the three frameworks
The axial load of 200 N was applied at the abutment region of right distal implants. Nonaxial load of 200 N was applied at an angle of 30° at the same site
Framework design on the implants.
Evaluation of the amount of stress distribution
The software gave quantitative values at different locations. Principal compressive and tensile strain along with von Mises stress was assessed in the peri-implant bone to evaluate the stress condition of the bone.
Evaluation of micromotion
Micromotion was computed as the relative displacement between two nodes (a node of bone side and implant side) of elements on the interface.
Strain distribution
The compressive and tensile strains were observed at the posterior implant and cantilever extension on axial and oblique loading. Maximum strain values were observed with the PEEK framework material. On the axial loading of the posterior implant, the tensile strain of Peek framework was 0.004888 and the compressive strain was 0.008001
Compressive strain at polyetheretherketone framework on axial loading of posterior implant.
von Mises stress distribution
von Mises stresses were higher on oblique loading of the titanium and PEEK framework materials in comparison to axial loading at cantilever length. The least amount of von Mises stresses was noted with the zirconia framework on axial and oblique loading of 21.0644 Mpa and 20.137Mpa, respectively
von Mises stress at zirconia framework on oblique loading of posterior implant.
Micromotion/displacement
The maximum micromotion was observed in the PEEK framework on axial loading at the posterior implant
Micromotion at polyetheretherketone framework on axial loading of posterior implant.
In the present study, stress distribution and micromotion at bone–implant interface using different framework materials in “All-on-four” concept in the edentulous mandible were analyzed.
Studies have shown the occlusal masticatory forces in the posterior region of the tooth to be around 220N.
The “All-on-four concept” eliminates the need for augmentation procedure in many cases. The angulation of distal implants also opens a window for placement of longer implants, thus enhancing the load distribution.
The knowledge of stress distribution around the implant–bone interface is crucial for its long-term stability. The stresses transferred onto the bone from different framework materials are divergent due to the variation in its Young's modulus of elasticity.
In the present study, principal tensile and compressive strains were observed at the crestal cortical bone. The length of the cantilever influenced the strain values. A study conducted by Horita et al. compared cantilever and non-cantilever loading in immediate implant placement. They found 45.3%–52.5% reduction in peak compressive strains on non-cantilever loading.
In the present study, at distal implant axial loading, the highest von Mises stress of 107.589 Mpa was noted with PEEK framework. However, the stress values were lower on the distal implant oblique loading of the three framework materials. The least stress of 20.137 Mpa was noted at zirconia framework. Thus, the framework material exhibits the ability to absorb stresses and distribute the minimal load to the adjacent bone. Overloading of the cortical bone occurs beyond 170Mpa. The value observed here is below that considered pathologic to the bone. Resorption due to excessive loading of adjacent bone could be avoided by the use of zirconia framework. The use of rigid framework material clinically would thus prevent failure of the implant support system.
The increasing anteroposterior spread of implants shortens the length of cantilever extension. The cantilever length affects the stress distribution of the underlying bone and the bone–implant interface.
Excessive micromotion at the implant–bone interface can have deleterious effects on the implant system. In this study, during axial loading, the least amount of displacement was observed at the zirconia framework. The titanium and PEEK materials showed similar values on axial loading. On oblique loading of the distal implant site, all the framework materials showed similar values, and the displacement was less than axial loading. In cantilever loading, maximum micromotion was observed on axial load than oblique load. Thus, according to the present study, it can be inferred that the overall loading angle and the type of loading affected the micromotion of the implant. The values did not differ significantly for cantilever and noncantilever loading. These results however differ from the study where the smallest micromotion was seen at noncantilever loading (<1/3
rd) compared to cantilever loading.
Prior studies suggest that lower Young's modulus of a framework material results in greater bending forces at distal implants and larger bending of the prosthesis under functional loading.
One of the limitations in this study includes the materials being assumed to have isotropic linear elasticity and inhomogeneity for bones. Although not seen in a clinical scenario, these are inherent in finite element studies due to limitations in biologic simulation. The study was also conducted under unilateral static loading. The framework materials may behave differently under cyclic loading such as those occurring during chewing movements. Further clinical studies are necessary for relevance and acceptance of the findings seen in the present study.
Within the limitations of this study, it was concluded that:
The stress distribution pattern at implant–bone interface was influenced by the framework material used. Zirconia framework showed minimal stress distribution in all parameters The site of load application and the direction of forces in axial and oblique patterns influence the compressive and tensile strain. PEEK framework material had the highest strain values The length of cantilever increased the stress concentration in all three framework materials The framework material, loading site, and direction of forces influenced the stresses and displacement at the bone–implant interface.