This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
Background:
Nickel-titanium (NiTi) open coils are common tools in fixed orthodontic treatments to apply light and continuous forces. Their favorable mechanical properties and their relatively high cost have prompted many clinicians to decide to reuse them.
Aim:
The present study aimed to evaluate the effect of 10% hydrochloric acid (HCl) and 5.25% sodium hypochlorite (NaClO) solution and sterilization, on the unloading force of NiTi open-coils.
Materials and Methods:
This experimental in vitro study consisted of 160 NiTi open coils from two brands of Highland Metals and Masel were provided. Each group was randomly divided into four subgroups: control, three rounds of immersion, three rounds of autoclave sterilization, and three rounds of immersion and autoclaving. Each round of immersion process included 1 min in 10% HCl, followed by 5 min in 5.25% NaClO. Spring was compressed for 4 mm. Then, the unloading forces of each spring were recorded at 4, 3, 2, and 1 mm of compression, respectively. The data were analyzed with SPSS (P < 0.05), using one-way ANOVA and independent t-test.
Results:
One-way ANOVA showed a significant difference in the 2-mm compression of Group 1 and 3-mm compressions of Groups 1 and 2 compared to the control group in the Highland metals brand (P < 0.04, P < 0.014, and P < 0.007, respectively). There was no significant difference between the different compressed lengths and groups in the Masel brand.
Conclusion:
One-minute immersion in 10% HCL and 5-min immersion in 5.25% NaOCl, followed by autoclave sterilization, even after repeating the process three times, did not significantly change the force properties of NiTi open springs. Therefore, the re-use of these springs with the above-mentioned method as preparation can be recommended.
Nickel-titanium (NiTi) alloy has been used in orthodontics since the early 1970s.
1Due to the proper mechanical properties of this alloy, its use has gradually increased, and it has won great popularity in recent decades.
2The positive characteristic of this alloy is related to its low load deflection rate and, consequently, high flexibility and elasticity.
3
NiTi springs are a common tool in fixed orthodontic treatments to apply light and continuous forces. These springs have been manufactured in both open and closed forms.
4,
5Open coils are used to provide the space for impacted teeth, severely mal-aligned teeth, distalization of molars, and retraction of incisors.
6,
7NiTi springs apply the more appropriate force in a longer activation range compared to steel springs. Thus, the force applied is more physiologic.
5,
8
NiTi alloys are expensive.
9Their favorable mechanical properties and their relatively high cost have prompted many clinicians to decide to re-use them.
10Economic considerations are the most important cause of their reuse.
11,
12Their manufacturing process needs to use limited natural resources; therefore, their reuse could help correct this imbalance in the environment.
13,
14By reusing, costs are reduced by 30%–40% and nature is protected against the depletion of resources.
13
Reuse of orthodontic appliances is possible when, in addition to preventing the transmission of infections, the physical characteristics of the appliance do not significantly change after the initial application, cleaning, disinfection, and sterilization processes.
15
It is possible to reuse tools and appliances that retain their function after clinical use by observing some principles. This is also implemented in medicine and other dental fields. The reuse of some appliances has also been suggested in orthodontics. These include brackets,
16,
17,
18,
19bands,
20wires,
21,
22,
23,
24orthodontic mini-implants.
25,
26
A literature search did not bring up a study on the reuse of NiTi open coils. However, some studies have evaluated the reuse of closed coils. Momeni et al. showed that autoclaving could increase their strength; however, the increase was not clinically significant. They claimed that Niti close coils could be reused without a significant change in their strength.
27Geng et al. reported that thermocycling resulting from the daily use of foods and drinks might result in the loss of the force of NiTi close coils.
28Magno et al. showed that 6-month clinical use of NiTi close coils could result in a significant increase in stress-strain curve, a decrease in strength, and permanent deformation.
29
The results on close coils are contradictory and needed more researches. It should be noted that the manufacturing conditions and distribution of force due to deformation in open and closed coils are different. There are limited studies available on the reuse of NiTi open coils. Due to their favorable characteristics and high cost, reusing of these springs can be suggested. The present study is aimed to evaluate the effect of the use of 10% hydrochloric acid (HCL) and 5.25% sodium hypochlorite (NaClO) solution to eliminate mineral and organic deposits on NiTi open coils, respectively, followed by sterilization, on unloading force of these springs.
Materials and Methods
Ethical approval was granted by the Torabinejad Dental Research Center, Isfahan University of Medical Sciences, Isfahan, Iran, with identification number 398835. This experimental in vitro study consists of 160 orthodontic NiTi open coils, 14 mm in length, including 80 from Highland Metals (Encinitas, California, USA) and 80 from Masel) San Jose, California, USA) brand. Each group was randomly divided into four subgroups. By setting a confidence level at 0.95, power at 0.80, and the standard deviation at 6, the sample size should be 20 samples for each group. Groups 0 were control, and Groups 1–3 underwent three rounds of immersion, three rounds of autoclave sterilization, and three rounds of immersion and autoclaving, respectively. Each round of immersion process included 1 min in 10% HCl (Household cleaner solution, Golrang Industrial Group, Tehran, Iran), followed by 10 s irrigation with running water and followed by 5 min in 5.25% NaClO (Bleaching solution, Golrang Industrial Group, Tehran, Iran). Each sample was immersed in a 20 mL fresh solution separately. The autoclaving process included 10 min in 121°C temperature and +1 bar pressure (Gemini BV, Apeldoorn, Netherland).
After this step, the first researcher coded the samples and the second researcher measured the spring unloading forces using an electromechanical universal testing machine (K-21046, Water + Bai, Switzerland) at room temperature
Figure 1. Spring was compressed up to 4 mm. Then, the unloading forces of each spring were recorded at 4, 3, 2, and 1 mm of compression, respectively. The data after decoding were analyzed with SPSS (version 22) SPSS (version 22) (SPSSInc., Chicago, IL,USA) (P < 0.05), using one-way ANOVA and independent t-test. The level of significance was set at 0.05. The normality was analyzed using Kolmogorov–Smirnov test.
Measuring unloading force of each coil.
Figure 1
Results
Kolmogorov–Smirnov test showed the normal distribution of data. One-way ANOVA analysis was carried out to compare the Groups at 1, 2, 3, and 4 mm of activation separately. The mean levels of force at different compressions are shown in
Table 1.{Table 1}
In the Highland metals brand, there were no significant differences between groups except in the 2-mm compression of Group 1 and 3-mm compressions of Groups 1 and 2 compared to the control group. (P < 0.04, P < 0.014, and P < 0.007, respectively) There was no significant difference between the different groups in the Masel brand. The control groups of the two brands compared with independent t-test. Masel springs had higher unloading forces than Highlands springs at 1 and 2 mm compression (P < 0.001, P = 0.017) but at 3 and 4 mm compression there were no significant differences (P < 0.066, P = 0.972).
Discussion
The most important rationale of medical and dental instruments reuse is economic issues.
11,
12The removal of organic and mineral debris as cleaning is necessary before sterilization. In this study, immersion in 10% HCL, followed by immersion in 5.25% NaOCl, was proposed as a method for cleaning NiTi open coils before autoclave sterilization. They are known as household cleaner and bleaching solution, and they are inexpensive and available. HCL can remove mineral debris.
16NaOCl can resolve organic debris and is a common and powerful disinfectant
25,
26Autoclave sterilization is carried out as a routine process of sterilization. Cleaning and sterilization processes must not have an adverse effect on the physical properties of equipment if they are reused.
30,
31
In this study, two different brands of NiTi open coils were used to help better the generalization of the results. Measuring unloading forces at 1–4-mm compression is done to simulate routine clinical use of these coils.
Three times immersion did not show significant changes in force characteristics of the Masel coils; however, the immersion procedure with Highland Metals coils at 2 and 3 mm compression resulted in significant changes, statistically, in force characteristics up to 12.6 and 20 cN, respectively, which is not significant clinically.
Three rounds of autoclave sterilization in the Masel group did not result in significant changes in force levels. However, concerning the Highland Metals coils, a decrease of 21.54 cN was recorded in force only at 3 mm compression, with no significant differences at other points. This amount of decrease in force is not significant clinically. Momeni et al.
27found that even though autoclave sterilization can increase the force levels of Ni-Ti close coils but this is not clinically significant, and it is possible to recycle NiTi closed-coil springs without a considerable reduction in their force levels.
No studies were found about the effect of reuse on NiTi open coils. However, several studies have investigated the effects of clinical use and sterilization on NiTi close coils and NiTi wires. Several studies have supported the reuse of NiTi close coils.
2,
27,
32,
33
Crotty examined the effect of autoclave on tensile properties of NiTi wires and reported no significant changes in the tensile properties of NiTi wires sterilized in an autoclave,
34which were consistent with Lee's findings.
35On the other hand, Bavikati et al. showed that autoclave sterilization of NiTi wires and clinical reuse of the wires could reduce the elasticity of the NiTi wires and affect their surface topography.
36Alavi also reported the changes in the mechanical properties of NiTi wires after autoclaving.
9There was no significant decrease in the force of any of the brands by concomitant autoclaving and immersion for three rounds. No similar study was found for comparison concerning the effect of the cleaning/sterilization process on the spring force.
Three times repeating the process of immersion and sterilization in this study were used to investigate the possibility of reuse for several times. By accepting the lack of the effect of the above process three times on the spring properties, the results could be generalized with greater confidence to less repetitions of the reuse.
Conclusion
One-minute immersion in 10% HCL and 5-min immersion in 5.25% NaOCl, followed by autoclave sterilization, even after repeating the process three times, did not significantly change the force properties of NiTi open springs. Therefore, the reuse of these coils with the above-mentioned method can be recommended.
Financial support and sponsorship
Nil.
Conflicts of interest
The authors of this manuscript declare that they have no conflicts of interest, real or perceived, financial or non-financial in this article.
HanSQuickDCNickel-titanium spring properties in a simulated oral environment6772WichelhausABrauchliLBallJMertmannMMechanical behavior and clinical application of nickel-titanium closed-coil springs under different stress levels and mechanical loading cycles6718KapilaSReichholdGWAndersonRSWatanabeLGEffects of clinical recycling on mechanical properties of nickel-titanium alloy wires42835SamuelsRHPeakJDUse of nickel titanium closed-coil springs to align unerupted teeth: A case report3538von FraunhoferJABondsPWJohnsonBEForce generation by orthodontic coil springs1458BrauchliLMSennCBallJWichelhausAForce levels of 23 nickel-titanium open-coil springs in compression testing6015PetrovADavidyanAGoonewardeneMSIncisor retraction with a compressed open-coil spring653SantoroMNicolayOFCangialosiTJPseudoelasticity and thermoelasticity of nickel-titanium alloys: A clinically oriented review.Part II: Deactivation forces594603AlaviSRGhorbaniAAEffects of steam and dry-heat sterilization on bending properties of NiTi wires1238MayhewMJKusyRPEffects of sterilization on the mechanical properties and the surface topography of nickel-titanium arch wires2326ManatakisDKGeorgopoulosNReducing the cost of laparoscopy: Reusable versus disposable laparoscopic instruments14UngerSLAssessing the environmental, human health, and economic impacts of reprocessed medical devices in a Phoenix hospital's supply chain19952003ThopegowdaNShenoyKShankarnarayanaRKukkilaJVaddyaSGinjupalliKRecycling of Materials used in Dentistry with Reference to its Economical and Environmental Aspects1405GrantcharovPAhmedSWacKRivasHReprocessing and reuse of single-use medical devices: Perceptions and concerns of relevant stakeholders toward current practices537MontgomerySRDisposable versus reusable blood pressure cuffs: A nursing led initiative13BahnasiFIAbd-RahmanANAbu-HassanMIEffects of recycling and bonding agent application on bond strength of stainless steel orthodontic bracketse197202HanRQYangKJiLFLingCAnalysis of shear bond strength and morphology of Er: YAG laser-recycled ceramic orthodontic brackets16YassaeiSAghiliHHosseinzadeh FirouzabadiAMeshkaniHEffect of Er: YAG laser and sandblasting in recycling of ceramic brackets1721KachoeiMMohammadiAEsmaili MoghaddamMRikhtegaranSPourghazneinMShiraziSComparison of multiple rebond shear strengths of debonded brackets after preparation with sandblasting and CO2 laser14854FulfordMRIrelandAJMainBGDecontamination of tried-in orthodontic molar bands6212OshaghMHematiyanMRMohandesYOshaghMRPishbinLAutoclaving and clinical recycling: Effects on mechanical properties of orthodontic wires63842PernierCGrosgogeatBPonsonnetLBenayGLissacMInfluence of autoclave sterilization on the surface parameters and mechanical properties of six orthodontic wires7281KhierSEBrantleyWAFournelleRAStructure and mechanical properties of as-received and heat-treated stainless steel orthodontic wires20612RamazanzadehBAAhrariFSabzevariBZebarjadSMAhrariAEffects of a simulated oral environment and sterilization on load-deflection properties of superelastic nickel titanium-based orthodontic wires1321NoorollahianSAlaviSMonirifardMA processing method for orthodontic mini-screws reuse44751EstelitaSJansonGChiquetoKFerreiraESEffect of recycling protocol on mechanical strength of used mini-implants15Momeni DanaeiSOshaghMKhozaeiAEffect of recycling and autoclave sterilization on the unloading forces of NiTi closed-coil springs: An in vitro study18490GengHSuHWhitleyJLinFCXuXKoCCThe effect of orthodontic clinical use on the mechanical characteristics of nickel-titanium closed-coil springs80314MagnoAFMonini AdaCCapelaMVMartinsLPMartinsRPEffect of clinical use of nickel-titanium springs7682EckelmanMMosherMGonzalezAShermanJComparative life cycle assessment of disposable and reusable laryngeal mask airways106772SloanTFirst, do no harm a framework for evaluating new versus reprocessed medical devices191201VidoniGPerinettiGAntoniolliFCastaldoAContardoLCombined aging effects of strain and thermocycling on unload deflection modes of nickel-titanium closed-coil springs: An in vitro comparative study4517AlaviSHaerianAThe effects of aging process and preactivation on mechanical properties of nickel-titanium closed coil springs2314CrottyOPDaviesEHJonesSPThe effects of cross-infection control procedures on the tensile and flexural properties of superelastic nickel-titanium wires3741LeeSHChangYIEffects of recycling on the mechanical properties and the surface topography of nickel-titanium alloy wires65463BavikatiVNSingarajuGSMandavaPKillamsettySSNettamVKarnatiPKEvaluation of mechanical and physical properties of clinically used and recycled superelastic NiTi WiresZC3540