Microstructural, microchemical, and mechanical changes associated with the clinical reuse of two nickel–titanium endodontic instruments

Felipe Augusto Restrepo‑Restrepo, Viviana Andrea Holguín‑Vásquez, Syldana Julieth Cañas‑Jiménez, Paula Andrea Villa‑Machado, Sara Ochoa‑Soto, Claudia Patricia Ossa‑Orozco, Sergio Iván Tobón‑Arroyave

Abstract


Background: Nickel–titanium (NiTi) instruments have represented a great technological
development that enabled endodontists conforming irregular‑shaped root canals. Notwithstanding,
the repeated use of these instruments may lead to the fracture without any prior visible warning
signs. This study aimed to evaluate how multiple clinical instrumentation/sterilization cycles of
two NiTi mechanized instruments can affect their microstructural, microchemical, and mechanical
characteristics.
Materials and Methods: In this observational descriptive study, a total of 140 NiTi instruments,
70 ProTaper Gold® (PTG) and 70 WaveOne Gold® (WOG) were analyzed. For each brand system,
instruments were evaluated in the as‑received condition (n = 10) and after one (n = 20), two (n = 20),
and three (n = 20) instrumentation/sterilization cycles. Intraoperative instrumentation parameters
were recorded for all used instruments. Afterward, the files were examined using scanning electron
microscopy and energy‑dispersive X‑ray microanalysis. All of the instruments were tensile‑fatigue
tested until rupture in order to calculate the mechanical tensile strength and the maximum elongation
percentage for the samples. Statistical analysis was completed using Chi‑square, Kruskal–Wallis
H‑, or Mann–Whitney U‑tests with a statistical significance set at P < 0.05.
Results: Significant increasing changes in surface topography (P < 0.05, Chi‑square test) and chemical
composition (P < 0.05, Kruskal–Wallis H‑test) in both brand systems through instrumentation/
sterilization cycles were detected. In addition, values of mechanical tensile strength and maximum
elongation percentage increased significantly through instrumentation/sterilization cycles in the
PTG group, whereas only the median values of mechanical tensile strength increased significantly
in the WOG group (all P < 0.01, Kruskal–Wallis H‑test).
Conclusion: Although multiple instrumentation/sterilization cycles may render NiTi instruments
more flexible and fatigue resistant, the significant changes detected in their surface topography
and chemical composition should preclude their repeated clinical use in the routine endodontic
practice as prevention for breakage.
Key Words: Electron probe microanalysis, endodontics, instrumentation, scanning electron microscopy, tensile strength