This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
Background:
This study presented a cleaning method for orthodontic NiTi-closed coils and evaluated its effect on the force characteristics of these coils.
Materials and Methods:
In this in vitro study, 160 orthodontic NiTi-closed coils, 160 orthodontic NiTi-closed coils (9 mm) from two brands G and H and DB, randomly assigned into four groups. Group 1: Three times of immersion in the 10% hydrochloric acid (HCI) solution for 1 min followed by the immersion in 5.25% sodium hypochlorite solution for 5 min. Group 2: Three times autoclave sterilization. Group 3: Three times of immersion in 10% HCI solution for 1 min, followed by the immersion in 5.25% sodium hypochlorite solution for 5 min and autoclave sterilization. Group 4: No immersion and no sterilization. The forces of coils were measured at 25, 23, 21, 19, and 17 mm extension. To measure the force characteristics of coils, an electromechanical universal testing machine was used. For the statistical analysis, Kolmogorov–Smirnov, Kruskal–Wallis, and Mann–Whitney U-tests were used at 0.05 significance level.
Results:
In G and H coils, no significant differences between the mean forces of different groups were found in any extensions. In DB coils, only in 25 mm extension, there were no significant differences between the groups (P = 0.144). In 17-, 19-, and 21-mm extensions, significant differences (P < 0.001) were found like in 23 mm extension (P = 0.05). In both brands, more extension makes significant more forces statistically (P < 0.01). In all extensions, G and H coils have significant less forces (P < 0.001).
Conclusion:
Immersion in HCl, 10% for 1 min followed by the immersion in NaOCl, 5.25% for 5 min followed by autoclave sterilization, even three times repeating of this process, did not result in significant changes in the force characteristics of A-NiTi-closed coils.
NiTi-closed coils are the routine force sources in sliding mechanics for space closing. The low and consistent forces of them are their principal advantage.
1Their high cost is the noticeable disadvantage. Therefore, their reuse has been considered.
2Reuse of medical instruments, such as the delicate angiography catheters, has a long history.
3This can be accomplished only in cases in which the first use and preparation of them for their reuse do not result in significant clinical properties changes. Furthermore, reuse should be possible from an economic point of view and easily implemented in the clinic.
4The most important reason for reuse is economic considerations.
5,
6Other reasons are bioenvironmental considerations.
6,
7
Many dental instruments are reused, for example, steel crowns,
8bone drills,
9polishing cups,
10polishing disks,
11endodontic files,
12and some dental implant components such as healing abutments.
5In orthodontics, reuse of some instruments has been suggested for many years,
13for example, metallic
14and ceramic
15brackets, bands,
16,
17NiTi wires,
2,
18and orthodontic mini-implants.
4,
19
It is necessary to clean the medical instruments before the sterilization process.
20The aim of the cleaning is to eliminate or decrease the visible particles, including blood proteins and debris from the surface.
20The elimination of mineral and organic debris has a pivotal role in infection control, particularly in preparing them for reuse.
21During the cleaning process, it is necessary to gain access to all the internal and external surfaces of the instrument.
20The type of the instrument and its design might make the cleaning process difficult. Surface irregularities and porosities, sharp angles, closed ends, dead spaces, rough borders, lumens, long and thin axes, and fissured surfaces make it difficult to eliminate the debris and biofilms.
20Residual bacterial debris, biofilms, and their drying on the surfaces make it difficult to clean and sterilize them.
20The clinical process might be mechanical, chemical, or a combination of both techniques.
22Chemical techniques consist of the use of solutions that have specifically been labeled for cleaning medical instruments. In general, the manufacture of each dental instrument, which can be reused, should explain the correct technique for disinfecting or sterilizing the instrument.
23
Hydrochloric acid (HCI) is a clear liquid with a corrosive behavior, which is used in home as a cleaning agent in addition to its applications in different industries.
24The concentration of this agent for household applications is 10%‒12% and is used for different cleaning and disinfecting purposes.
25,
26In dentistry, this solution is used to remove the enamel stains,
27microabrasion to improve decalcification.
28This powerful acid can be used to eliminate the mineral debris. Sodium hypochlorite is an inexpensive-oxidizing agent and the most commonly used disinfecting agent. One of the main characteristics of sodium hypochlorite is its antibacterial properties due to its proteolytic potential and buffering capacity.
29This solution is capable of eliminating the organic tissues.
Limited studies have evaluated the reuse of NiTi-closed coils in orthodontics.
30,
31This study presented a novel-cleaning method and evaluated its effect and autoclave sterilization on the force characteristics of these coils.
Materials and Methods
This study is a blind-experimental study. 160 NiTi closed coils, with 9 mm length from two brands G and H (G and H Wire Company, 2165 Earlywood Drive, Franklin, IN, USA) and DB, (DBOrthodontics, Ryefeild Way, Silsden, West Yorkshire, UK) were used. Each brand is randomly assigned into four groups (n = 20). The inclusion criteria consisted of the absence of any visible and definitive macroscopic defects in the coils. The exclusion criteria were any visible and definitive macroscopic defects in the coils. The coils were mounted on the jigs fabricated with 0.9-mm steel wires, so that the internal distance of the coil eyelets was 25 mm
Figure 1. In each brand, the test groups underwent different processes as follows:
Jigs for mounting the coils; were fabricated with 0.9-mm steel wires so that the internal distance of the coil eyelets was 25 mm
Figure 1
Group 1: Three times immersion in the household cleaning solution for 1 min followed by the immersion in home bleach solution for 5 min
Group 2: Three times autoclave sterilization. (at 134°C and 32 psi)
Group 3: Three times of immersion in household disinfecting solution for 1 min, followed by the immersion in home bleach solution for 5 min and autoclave sterilization
Group 4: No immersion and no sterilization (new coils as control).
The household cleaning solution brand was Golrang (Golrang, Tehran, Iran) and consisted of 10% HCI. The home bleach solution consisted of 5.25% sodium hypochlorite solution from Golrang (Golrang, Tehran, Iran). The solutions were prepared fresh with a volume of 20 ml for each sample. After each immersion, the samples were rinsed under running water for 10 s, and then, the next immersion procedure was carried out.
To evaluate the effects of autoclave sterilization, the samples were placed in an open container and transferred to an autoclave (MOCOM; PRIMA, Australia). The coils were placed far from each to provide enough ventilation for a better sterilization. The process was done at 134°C and 32 psi.
For the blinding of measurements, after these steps, the samples were coded by the first researcher, and all the measurements were carried out by the second researcher.
To measure the force characteristics of coils, an electromechanical universal testing machine (K-21046, water + bai, Switzerland) was used. Two hooks were fabricated with 0.9-mm steel wire and placed on each side of the machine so that the distance between them was 25 mm. The first point for measuring the force (25 mm) was approximately equal to the distance between the hook of the canine tooth and that of the first molar tooth. Each coil was placed separately on this setup, and its force was measured. Then, the moving arm moved toward the other at speed of 5 mm/min
1and the unloading force of each spring was recorded at each 2 mm steps. Hence, the forces of coils were measured at 25, 23, 21, 19, and 17 mm extensions. The effective part of the coil, i. e., the part with an elastic capacity, was around 4 mm less than that reported for the coil.
37The effective length of coil were measured under a stereomicroscope with 8X zoom (Trinocular zoom Stereo Microscope, SMP-200, Supplied with Moticam 480 DigitalCamera, SP10.0224, Motic Instruments Inc., CA, USA)
Figure 2. All the tests were carried out under the same condition at the room temperature. One sample from each group underwent evaluations under the scanning electron microscope (SEM) with 50, 150, and ×1500 zoom for qualitative evaluation of surface changes
Figure 3. After decoding of the samples, data were analyzed with Kolmogorov–Smirnov, Kruskal–Wallis, and Mann–Whitney U tests with the SPSS software version 22 (SPSS Inc., Chicago, IL, USA). Significance level was set at 0.05.
Measuring the effective lengths of closed coils with stereomicroscope
Figure 2
SEM views
Figure 3
Results
The mean of force values in different extensions and effective coil lengths is presented in
Table 1and
Table 2, respectively. One-sample Kolmogorov–Smirnov test showed that only 16 from 48 groups had normal distribution. Hence, nonparametric tests were used. Kruskal–Wallis test was done to compare four different groups in 17, 19, 21, 23, and 25 mm coil extensions, separately.{Table 1}{Table 2}
In G and H coils, no significant difference between the mean forces of different groups was found in any extensions. In DB coils, only in 25 mm extension, there was no significant differences (P = 0.144) between the groups. In 17, 19, and 21 mm extensions significant differences (P < 0.001) were found like in 23 mm extension (P = 0.05). In both brands, more extension makes significant more forces statistically (P < 0.01). In all extensions, G and H coils have significant fewer forces (P < 0.001). The mean of coil lengths in both G and H (P = 0027) and DB (P < 0.001) brands had a significant difference between their groups.
In electron microscopic views, no significant changes were seen between different groups of both brands. The surface of v coils seems smoother.
Discussion
Reuse of instruments that are not damaged or do not undergo major changes in their mechanical properties results in a decrease in treatment costs. This can be great to manage of limited resources of healthcare, and finally, leading to increase in the number of patients benefiting from therapeutic services.
32,
33
Some studies have evaluated the effect of the clinical use and oral cavity conditions, including thermal and mechanical cycles, on NiTi-closed coils.
1,
30,
34,
35,
36,
37Most of them concluded that these factors do not have a significant clinical effect on the force levels of coils.
Before sterilization, cleaning is necessary.
38In the present study, immersion in 10% HCI, followed by 5.25% sodium hypochlorite, was proposed as the cleaning method for NiTi-closed coils. These solutions are available as inexpensive household cleaning and bleaching agents. HCI, as a strong acid, has a high capacity to eliminate calcium-based minerals. Sodium hypochlorite, apart from its capacity to eliminate organic debris, exhibits high-disinfecting properties. However, it does not damage titanium and its alloys. It is necessary to gain access to all the surfaces of instruments that are prepared for reuse to carry out an effective cleaning procedure. This is more difficult for small objects with specific shape such as coil springs. The separation of coil helices due to mounting on jig and liquidness of cleaning agents make better access to all the surfaces of coil. Good solubility of them in water makes removing them easy with washing with water.
Two different brands of NiTi-closed coils were used: G and H as an A-NiTi coil, and DB as an M-NiTi coil. Sterilization in an autoclave and repetition of the process for three times did not result in a significant change in force characteristics of G and H coils.
However, significant changes were observed in the force properties of DB coils. Momeni et al., too, reported a minor effect of sterilization in an autoclave on the force properties of these coils.
31Ramazanzadeh et al.
39reported that sterilization with autoclave resulted in a minor increase in the force levels of NiTi wires, which was not significant statistically. Smith et al.
40evaluated the effect of clinical use and different disinfection/sterilization protocols on three types of NiTi wires and concluded that there were no significant clinical differences between the wires. Alavi et al.
2reported that dry heat and autoclave sterilization resulted in a significant decrease in the forces of NiTi wires. Bavikati et al.
18reported significant decreases in tensile strength and surface roughness of NiTi wires due to clinical use and sterilization. It should be noted that the manufacturing conditions and force distributions due to the changes in their form are different between NiTi wires and coils. Therefore, caution should be exercised in extending the results of wires to coils.
Immersion in 10% HCI, followed by immersion in 5.25% sodium hypochlorite and repetition of the procedure for three times did not result in significant changes in the force characteristics of G and H and BD coils. Immersion in 10% HCI, followed by immersion in 5.25% sodium hypochlorite, and then, sterilization in an autoclave and repetition of this process for three times did not result in significant changes in the force characteristics of G and H coils. Therefore, cleaning and sterilization of A-NiTi-closed coils to reuse is possible. In relation to the effect of immersion in the disinfecting solution in the present study or the effect of cleaning alone or in association with sterilization in an autoclave, no similar study was found on orthodontic coils for the comparison. However, it has been reported that the immersion in 5.25% NaOCl for 24 h did not affect the surface characteristics and fracture resistance of NiTi files.
41
In both coil types, an increase in extension resulted in an increase in force levels, with higher rate in the DB coils, which might be explained by its different types of NiTi alloy (M-NiTi). The less steep force diagram of G and H coils is related to its super-elastic properties. No similar study was found on the effect of use or cleaning/sterilization procedure on the coil length for the comparisons. Comparison of the coils showed that at similar extension the mean force in the DB coil was significantly higher. This can be explained by M-NiTi alloy of DB coils and A-NiTi alloy of G and H ones.
The first point for measuring the force (25 mm) was approximately equal to the distance between the hook of the canine tooth and that of the first molar tooth. This distance has also been used in similar studies.
42,
43,
44The point of termination for force measurement was 17 mm, approximately the distance after the canine tooth reaches the second premolar tooth. In addition, the measurements were carried out at points 19, 21, and 23 mm to make the results more valid and make the force changes gradient clearer at different extensions. Force measurement in the unloading condition is more similar to the clinic.
1,
34
The means of the forces measured in new coils in the control groups were different from those reported by the manufacturers, consistent with the results of previous studies,
1,
45which might be attributed to the differences in the conditions and techniques of measurements.
After the measurement of forces, the changes of effective length of coils were assessed under a stereomicroscope with 8X zoom (Trinocular zoom Stereo Microscope, SMP 200, USA). The effective part of the coil, i.e., the part with an elastic capacity was around 4 mm less than that reported for the coil.
37In relation to G and H coil, there were no significant differences in coil lengths between the control and experimental groups, which might be attributed to the supper-elastic nature of its alloy. However, in relation to DB coils, there were significant differences in coil lengths between the control and experimental groups. Greater permanent deformity of M-NiTi alloy compared to A-NiTi alloy was predictable and has been reported in other studies, too.
46
Repetition of the immersion and sterilization process in the present study aimed to evaluate the possibility of several times reuse. Repetition of intervention has been carried out in other studies.
11,
47This provides more confidence to generalize the results to fewer times of reuse. SEM images did not reveal significant surface changes due to immersion and autoclave sterilization procedures
Figure 4. A study reported insignificant increase in surface roughness after autoclave
48and other reported an increase in the surface roughness of NiTi wires
18and NiTi endodontic files
49after repeated autoclave sterilization. A study reported 24 h immersion in 5.25% NaOCl has some effects on surface characteristics of NiTi endodontic files.
50
Comparison of force values (cN) in different extensions of the two coil brands. Group 1: Three times immersion; Group 2: Three times autoclave sterilization; Group 3: Three times immersion and three times autoclave sterilization; Group 4: Control
Figure 4
Conclusion
Three times immersion in household-cleaning liquid (HCl, 10%) for 1 min and immersion in household bleaching liquid (NaOCl, 5.25%) for 5 min did not result in significant changes in the force characteristics of A-NiTi closed coils
Three times autoclave sterilization did not result in significant changes in the force characteristics of A-NiTi-closed coils
Three times immersion in household-cleaning liquid (HCl, 10%) for 1 min followed by the immersion in household-bleaching liquid (NaOCl, 5.25%) for 5 min followed by autoclave sterilization, did not result in significant changes in the force characteristics of A-NiTi closed coils.
Hence, A-NiTi closed coils can be reused and this immersion protocol can be suggested as cleaning method before autoclave sterilization.
Financial support and sponsorship
Nil.
Conflicts of interest
The authors of this manuscript declare that they have no conflicts of interest, real or perceived, financial or nonfinancial in this article.
AlaviSHaerianAThe effects of aging process and preactivation on mechanical properties of nickel-titanium closed coil springs2314AlaviSRajiSHGhorbaniAAEffects of steam and dry-heat sterilization on bending properties of NiTi wires1238MalcheskyPSChamberlainVCScott-ConnerCSalisBWallaceCReprocessing of reusable medical devices14651NoorollahianSAlaviSRafieiEThe effect of multiple processing and re-use on orthodontic mini-screw torque values2437WadhwaniCSchonnenbaumTRAudiaFChungKHIn-Vitro study of the contamination remaining on used healing abutments after cleaning and sterilizing in dental practice106974ThopegowdaNShenoyKKiaakkar ShankarnarayanaRKukkilaJBhat VaddyaSGinjupalliKRecycling of Materials used in Dentistry with Reference to its Economical and Environmental Aspects; 2013.pRecycling of Materials used in Dentistry with Reference to its Economical and Environmental Aspects; 2013 p 140-5DuaneBHarfordSRamasubbuDStancliffeRPasdeki-ClewerELomaxREnvironmentally sustainable dentistry: A brief introduction to sustainable concepts within the dental practice2925DarshanVIndushekarKRSarafBGSheoranNSharmaBSardanaDA comparison of decontamination methods of tried-in preformed metal crowns: An in-vivo study53744AllanWWilliamsEDKerawalaCJEffects of repeated drill use on temperature of bone during preparation for osteosynthesis self-tapping screws3149HeintzeSDForjanicMPolishing performance of multiple-use silicone rubber-based polishing instruments with and without disinfection/sterilization28894TürkünLSEffect of re-use of a disposable micro-polisher on the surface of a microhybrid resin composite27982BrysonLMFernandez RivasDBoutsioukisCCleaning of used rotary nickel-titanium files in an ultrasonic bath by locally intensified acoustic cavitation45768BuchmanDJEffects of recycling on metallic direct-bond orthodontic brackets65468de Oliveira CorreiaAMde Souza MatosFPilli JóiasRde Mello RodeSCesarPFParanhosLRComparative assessment of different recycling methods of orthodontic brackets for clinical use10714MirhashemiAHHosseiniMHChiniforoushNSoudiAMoradiMShear Bond Strength of Rebonded Ceramic Brackets Using Four Different Methods of Adhesive Removal5462BensonPEDouglasCWDecontamination of orthodontic bands following size determination and cleaning1824DowsingPBensonPEMolar band re-use and decontamination: A survey of specialists307BavikatiVNSingarajuGSMandavaPKillamsettySSNettamVKarnatiPKEvaluation of mechanical and physical properties of clinically used and recycled superelastic NiTi wiresZC3540NoorollahianSAlaviSMonirifardMA processing method for orthodontic mini-screws reuse44751SonntagDPetersOAEffect of prion decontamination protocols on nickel-titanium rotary surfaces4426TeterJZenilmanMEWachter P; Johns Hopkins MedicineCleaningDisinfection and Sterilization ClinicalCommunityAssessment of endoscope reprocessing using peer-to-peer assessment through a clinical community26570BoyceJMModern technologies for improving cleaning and disinfection of environmental surfaces in hospitals10ParkJHOlivares-NavarreteRBaierREMeyerAETannenbaumRBoyanBDEffect of cleaning and sterilization on titanium implant surface properties and cellular response196675Available from:https://pubchemncbincbinlmnihgov/compound/hydrochloric [Last visited in 2019 May 05 and Last accessed in 2022 Nov 30]Available from:http://www3EpaEpagov/airtoxics/hlthef/hydrochlhtml [Last visited in 2019 May 05 and Last accessed in 2022 Nov 30]Available from:http://wwwnlmnlmnihgov/medlineplus/ency/article/002498htm [Last visited in 2019 May 05 and Last accessed in 2022 Nov 30]BassirMMBagheriGComparison between phosphoric acid and hydrochloric acid in microabrasion technique for the treatment of dental fluorosis414SamuelsRHRudgeSJMairLHA clinical study of space closure with nickel-titanium closed coil springs and an elastic module739Frough-ReyhaniMGhasemiNSoroush-BarhaghiMAminiMGholizadehYAntimicrobial efficacy of different concentration of sodium hypochlorite on the biofilm of Enterococcus faecalis at different stages of developmente4804WichelhausABrauchliLBallJMertmannMMechanical behavior and clinical application of nickel-titanium closed-coil springs under different stress levels and mechanical loading cycles6718Momeni DanaeiShOshaghMKhozaeiAEffect of recycling and autoclave sterilization on the unloading forces of NiTi closed-coil springs: An In vitro study18490HuangTHYenCCKaoCTComparison of ion release from new and recycled orthodontic brackets6875ReimannSRewariAKeiligLWiduFJägerABourauelCMaterial testing of reconditioned orthodontic brackets45466VidoniGPerinettiGAntoniolliFCastaldoAContardoLCombined aging effects of strain and thermocycling on unload deflection modes of nickel-titanium closed-coil springs: An in-vitro comparative study4517Espinar-EscalonaELlamas-CarrerasJMBarrera-MoraJMAbalos-LasbrucciCGil-MurFJEffect of temperature on the orthodontic clinical applications of NiTi closed-coil springse7214BezroukABalskyLSmutnyMSelke KrulichovaIZahoraJHanusJThermomechanical properties of nickel-titanium closed-coil springs and their implications for clinical practice31927MaganziniALWongAMAhmedMKForces of various nickel titanium closed coil springs1827ParashosPLinsuwanontPMesserHHA cleaning protocol for rotary nickel-titanium endodontic instruments207RamazanzadehBAAhrariFSabzevariBZebarjadSMAhrariAEffects of a simulated oral environment and sterilization on load-deflection properties of superelastic nickel titanium-based orthodontic wires1321SmithGAvon FraunhoferJACaseyGRThe effect of clinical use and sterilization on selected orthodontic arch wires1539Tanomaru-FilhoMGalletti EspirCCarolina VençãoAMacedo-SerranoNCamilo-PintoJGuerreiro-TanomaruJCyclic fatigue resistance of heat-treated nickel-titanium instruments3127NattrassCIrelandAJSherriffMAn investigation into the placement of force delivery systems and the initial forces applied by clinicians during space closure12731OshaghMHematiyanMRMohandesYOshaghMRPishbinLAutoclaving and clinical recycling: Effects on mechanical properties of orthodontic wires63842SantosACTortamanoANaccaratoSRDominguez-RodriguezGCVigoritoJWAn in vitro comparison of the force decay generated by different commercially available elastomeric chains and NiTi closed coil springs517MelsenBToppLFMelsenHMTerpSForce system developed from closed coil springs5319SonAYLimSHLoad-deflection characteristics and plastic deformation of NiTi closed coil springs3109NairASTilakchandMNaikBDThe effect of multiple autoclave cycles on the surface of rotary nickel-titanium endodontic files: An in vitro atomic force microscopy investigation21822LeeSHChangYIEffects of recycling on the mechanical properties and the surface topography of nickel-titanium alloy wires65463ValoisCRSilvaLPAzevedoRBMultiple autoclave cycles affect the surface of rotary nickel-titanium files: An atomic force microscopy study85962de Castro MartinsRBahiaMGBuonoVTHorizonteBThe effect of sodium hypochlorite on the surface characteristics and fatigue resistance of ProFile nickel-titanium instrumentse99105