?
DRJDent Res JDental Research JournalDental Research Journal
This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
Background:
The present study was performed to evaluate the effect of three types of mouthwash (orthokin, oral B and chlorhexidine [CHX]) on releasing of aluminum (Al), Titanium (Ti) and Vanadium (V) ions from titanium mini-implants (TMIs).
Materials and Methods:
In this in vitro, experimental study, a total of 40 TMIs were divided equally into four groups (10 TMI in each group) and then were immersed into Orthokin, Oral B, CHX, and artificial saliva, as a control. The experiments were performed for 21 days as following groups 1–7 days, 8–14 days, and 15–21 days. The inductively coupled plasma-optical emission spectrometry method was used to assess releasing metal ions after immersion in the storage media. In addition, before and after each experiment, the corrosion of TMIs was assessed using a scanning electron microscope (SEM). All results were analyzed using Kruskal–Wallis, followed by Bonferroni-adjusted Mann–Whitney U-test at 0.05 level of significance.
Results:
Our data showed that the maximum concentration of released Al was in the 1
stweek of exposure to Orthokin and Oral B (202.3 ± 68.5 and 72.3 ± 15.2 μg/L, respectively). Oral B exposure of TMI also caused to releasing of Ti to 128.1 ± 42.5, 54 ± 19.4 and 22 ± 6 μg/L for 1–7 days and 8–14 days and 15–21 days, respectively. Orthokin and CHX also induced the release of Ti more than artificial saliva (P < 0.05). In addition, there was no significant statistical difference between any types of mouthwashes and artificial saliva in releasing V. The results of SEM images also confirmed the corrosion effects of mouthwashes.
Conclusion:
The factors of exposure time and mouthwash type influenced the pattern of releasing Al and Ti as well as corrosion level.
Corrosionmouthwashesscanning electron microscopytitanium alloy tial6 v4Introduction
Mini-implants (MIs) as the most commonly used temporary anchorage devices have been paid lots of attention. This type of treatment has many advantages, including inexpensiveness as well as simplicity in both placement and removal. MIs also have reduced the need for compliance of the patients without surgery. In addition, MIs have the ability to control appropriate anchorage in three spatial plans providing clinicians to improve orthodontic treatment. However, compared to endosseous implants, MIs have a reduced success rate, dependent on some conditions ranging from 70% to >95%. MIs failure has been associated with some factors such as age, maintenance, bone density, root proximity, the screw design, the placement method, and chemical composition.
1Furthermore, peri-implant tissue inflammation has been associated with loosening of the miniscrews in clinical studies.
2,
3The causes of the inflammation are multifactorial and include the level of patient health, surrounding tissue type, and design of the mini-screw head. Recently, corrosion has been considered as one of the factors related to peri-implantitis of TMIs with the release of titanium (Ti) ions. It is noted that macrophages surrounding failed dental implants have been found to be loaded with Ti from corrosion.
4
The chemical composition of TMIs also is of interest because it has been shown that corrosion of bracket, archwire, and other metallic fixed orthodontic appliances can occur in the oral environment.
5,
6,
7The typical composition of orthodontic MIs is Ti alloy, but certain amounts of aluminum (Al) and Vanadium (V) to provide stability, low toxicity, and resistance are added. However, these added ions make the alloy more susceptible to corrosion.
8
Practitioners recommend the daily use of fluoride mouthwashes to reduce the risk of development of white spots around orthodontic brackets during orthodontic treatment. However, some contents of mouthwashes such as fluoride ions, as well as the lower mouthwash pH, have been reported to enhance release TMIs ions causing corrosion and discoloration.
9,
10Chlorhexidine (CHX) mouthwash is recommended also to prevent inflammation around the MI.
11,
12These agents could lead to the corrosion of the implant head, which is exposed to the oral cavity.
Corrosion and releasing of metal ions from TMIs into biologic fluids appear to have a wide variety of harmful effects. Metal released from dental alloys has been reported to cause severe biological alterations. For instance, Natarajan et al. showed that the oral mucosal cells of healthy patients undergoing orthodontic treatment were damaged.
13The released ions can cause oxidative stress damaging to human cells. Oxidative stress seems to affect proteins, lipid, and nucleic acid, which consequently results in tissue damages and cancer progression.
14,
15For instance, The released Nickel is appeared to has high uptake, transport, distribution, and retention at a cellular level.
16,
17Cell culture studies have shown that Ti ions alter the phenotype and function of T-lymphocytes, induce the differentiation of monocytes into active osteoclasts and increase the expression of cytokines in osteoclasts.
18
The current study is aimed to assess the levels of metal ions released from orthodontic TMIs in the presence of three types of OralB, Orthokin, and CHX mouthwashes. Indeed, these findings should guide practitioners to decide which mouthwash to prescribe for each patient.
Materials and Methods
By the ethics and research committee of, Isfahan University of Medical Sciences (No.398893). Totally, 40 TMI (Jeil Medical, Seoul, Korea) were used for this in vitro, experimental study. The TMIs were divided randomly into four equal groups and immersed in artificial saliva (Kin Hidrat spray; Spain) and Oral B (Procter & Gamble, Weybridge, United Kingdom), Orthokin (Ortho-kin, Kin, Spain) and CHX (Iran Najo, Tehran, Iran) mouthwashes. These mouthwashes were chosen due to their commercial availability. MIs in each group were stored individually in a tube and incubated at 37°C for 21 days, which were divided into three separated immersion intervals as follows: 1–7 days, 8–14 days, and 14–21 days.
Inductively coupled plasma-optical emission spectrometry (ICP OES) was generally applied for the determination of metals in body fluids and other liquids. For this purpose, multi-elemental calibrants were prepared from different standard solutions of the individual elements. Calibration curves were also prepared with matrix-matched calibration standards. For this experiment, the samples were prepared as described previously.
19Emission intensities of samples were measured at 308.215 nm, 334.936 nm, and 292.399 nm for Al, Ti and V, respectively. The samples were analyzed using ICP Perkin-Elmir 7300DV (USA) in Axial mode.
Scanning electron microscope observations
Qualitative analysis of the topography of the head of MIs was performed before and after their exposure to experimental media using a scanning electron microscope [SEM]) INCAx-sight, England) at ×5000 magnification. One representative sample of each of the four groups randomly were analyzed.
Statistical analysis
Statistical analysis of the results was performed using SPSS 22 (SPSS Inc., Chicago, Illinois, USA). Normal distribution of data was assessed by the nonparametric Kolmogorov–Smirnov test. Since some variables presented abnormal distribution, nonparametric tests were employed (Kruskal–Wallis, followed by Bonferroni-adjusted Mann–Whitney U-test). All results were analyzed at a significance level of 0.05.
Results
The concentration of the ions released in three experiment groups and control in a different time is shown in
Table 1,
Table 2,
Table 3. As analyzed by the Kolmogorov–Smirnov test, all ions had nonnormal distribution; therefore, the nonparametric test was used. As reported in
Figure 1, the maximum concentration of Al was released in 1–7 days in the presence of Orthokin (202.3 ± 68.5 μg/L, P < 0.001). The second higher concentration of Al was indicated in the OralB group in a time of 1–7 days (72.3 ± 15.2 μg/L, P < 0.001). As indicated in
Figure 1, there was no relationship between time of exposure to mouthwashes and released ion concentration. In addition, Oral B exposure of TMI caused to release of Ti to 128.1 ± 42.5 μg/L, P < 0.001 and 54 ± 19.4 μg/L, P < 0.001 and 22 ± 6 μg/L, P < 0.05 for 1–7 days and 8–14 days and 15–21 days, respectively. Orthokin and CHX also caused to release of Ti more than artificial saliva at various times (P < 0.05). Similar to Al, there was no relationship between time of exposure to mouthwashes and released Ti concentration
Figure 2. Furthermore, V determination of samples showed that there was no significant statistical difference between any types of mouthwashes and artificial saliva at various times
Figure 3.{Table 1}{Table 2}{Table 3}
Released concentration of Al in response to three types of mouthwashes (orthokin, oral B and chlorhexidine) in different immersion times. *, ** and *** were stands for statistical significances of
P< 0.05,
P< 0.01 and
P< 0.001.
Figure 1
Released concentration of Ti in response to three types of mouthwashes (orthokin, oral B and chlorhexidine) in different immersion times. *, ** and *** were stands for statistical significances of
P< 0.05,
P< 0.01 and
P< 0.001.
Figure 2
Released concentration of V in response to three types of mouthwashes (orthokin, oral B and chlorhexidine) in different immersion times. There is no statistically significances difference between each groups.
Figure 3
Before the immersion of TMI in mouthwash and artificial saliva, the SEM analysis was performed to detect probably manufacture defects
Figure 4. After the indicated time of immersion period (1–7, 8–14, and 15–21 days) in the four testing solutions of Oral B, Orthokin, CHX, and artificial saliva SEM analysis were again performed. Representative SEM image of TMIs at various times is shown in
Figure 5. Dark spots, pits, and cracks and some corrosion products were observed on the surfaces of TMIs after immersion in mouthwashes. The results of SEM images showed a similar pattern of releasing ions. Orthokin caused corrosion more than Oral B, while CHX treatment resulted in minimal corrosion. Loss of gloss and increased corrosion in the surface of TMI by increasing the time of exposure were a significant result.
High-magnification SEM photomicrographs of titanium mini-implants without immersion (×5000).
Figure 4
High-magnification scanning electron microscope photomicrographs of titanium mini-implants after immersion in different solutions (×5000).
Figure 5
Discussion
Today during orthodontic treatment, practitioners recommend that their patients use mouthwashes daily to reduce the risk of dental caries. They also suggest that after using mouthwash, the patient should not eat, drink, and rinse; therefore, the chemical agents of mouthwash are exposed to TMIs for a long period that may lead to TMI corrosion.
20It is difficult to determine the exact contact duration between TMIs and mouthwash, but there is an assumption that in each 30–60 s, the mouthwash is present for 2–4 h. Orthodontic treatment is usually done in a long period, for about 1–2 years; therefore, 21 days is the arbitrary total duration assumption to immerse the TMIs into the mouthwashes.
Although, this is generally accepted that due to the stability of the metal oxide layer, Al, Ti, and V alloys are extremely resistant against corrosion. However, due to the chemical and mechanical properties of the oral cavity as well as microbial flora, enzymatic reaction, and thermal changes, some metal ions can be released from TMIs.
To date, the release of elements from implants has been indicated in different ways such as dental Atomic Absorption Spectroscopy,
21,
22ICP-OES
23and ICP-Mass Spectrometry (MS).
24Among such methods, the best limits of detection, as well as higher sensitivity and specificity, are provided by ICP-MS.
25Due to the lower power of detection, ICP-OES is unable to determine the elemental background. However, this simple method is satisfying in the determination of Al, Ti, and V.
26In addition, corrosion assessment of TMI is commonly evaluated by SEM This experiment was also validated by many studies.
27,
28,
29
In the present study, three types of mouthwash that commonly are recommended for patients were evaluated. Our findings showed that Al has a releasing peak in a time of 1–7 days in the presence of Orthokin and OralB. The World Health Organization (WHO) recommended that the aluminum tolerable daily intake is 1 mg/kg body weight/day.
30Due to the low concentration of Al (202.3 ± 68.5 and 72.3 ± 15.2 μg/L, for Orthokin and OralB, respectively) in comparison to approved daily intake, there was no matter for using Orthokin and OralB. Similarly, our data also indicated that Oral B exposure of TMI caused to release of Ti to 128.1 ± 42.5, 54 ± 19.4 and 22 ± 6 μg/L for 1–7 days, 8–14 days, and 15–21 days, respectively. Because of low Ti toxicity, it has no toxicological reference values. On the other hand, V determination of samples exhibited that there was no statistically significant difference between any mouthwashes and artificial saliva. Among released ions, the lowest amount was for V. similarly Abboodi and Al-Dabagh showed that the least amount of release was V.
27This phenomenon may be associated with the composition and behavior of the passive oxide surface layer, which is composed mainly of TiO
2, with small amounts of Al
2O
3, hydroxylic groups, and water. As V is not present in the superficial oxide layer of Ti-6Al-4V, Ti and Al are the metal ions most likely to be released from the Ti-6Al-4V surface.
31
The released pattern of ions is more likely to be associated with fluoride concentration of mouthwashes. The concentration of fluoride was 500 ppm, 250 ppm, and 0 ppm for Orthokin, OralB and CHX, respectively. Because CHX that has no fluoride was lower in releasing ions. While the most ions were released by Orthokin that has a higher concentration of fluoride. This is consistent with previous studies that show in acidic fluoridated solutions, the F − combines with H + to form hydrofluoric acid (HF), which is capable of destroying the oxide layer on Ti and its alloys. Soluble Ti-fluoride compounds form, leading to the dissolution of the metal.
32Kang et al. also evaluated the effects of acetic NaF solutions on Ti and Ti alloy brackets and noticed that the HF formed in acetic NaF solutions induced corrosion on the surface of Ti-based orthodontic brackets and with the increasing concentration of NaF the level of corrosion increased.
33
The concentration of all detected ions in response to all mouthwashes peak at day 7 and decrease in the next weeks of study. Such a reduction in the rate of ions release could be explained by the normal capacity of the passivation layer of the alloy to form a protective stable oxide film that inhibits the corrosion process in various environments; thus, the rate of corrosion will be decreased.
34Similarly, some of the previous studies have revealed that metal ions released from fixed orthodontic appliances have a peak of releasing on day 7.
25,
35,
36
On the other hand, the obtained results of released ions were also correlated with corrosion study, as indicated by SEM analysis.
Taken to gather our finding confirmed that although mouthwashes treatment of TMI resulted in releasing of Al and Ti in some period, according to the WHO standards, all types of mouthwashes are safe to be used. These data highlight that this type of MIs and CHX mouthwash is suitable for orthodontic purposes and meet the standard norms regarding the ingredient of CHX, and resistance of TMI against corrosion. However, other mouthwashes due to releasing the peak of Al and Ti at the time of 1–7 days must be paid attention.
Similarly, Pavlic et al. assessed corrosion of MIs in response to chlorhexidine and probiotic agents. They reported that for patients undergoing orthodontic TMIs, CHX could be recommended for oral-hygiene maintenance.
37Furthermore, Mandsaurwala, Mohammed et al. indicated that CHX-containing (HEXIDINE) mouthwash could be offered for patients who have orthodontic appliances rather than Na + alcohol-containing mouthwash (LISTERINE). They explained that greenish discoloration on the CHX sample occurs after 24 h, which constituted a type of passive surface.
38Although Danaei et al. reported that ion releasing from stainless steel brackets in response to deionized water was significantly higher than that for OralB, CHX, and Persica mouthwashes. They indicated that Oral B and Persica mouthwashes might be better options than CHX for orthodontic patients.
5The difference between these results and ours could be due to the material of brackets that they used. Stainless steel is vulnerable against CHX, as Pavlic et al. also confirmed.
37
Whether the ion released from TMIs (despite the safety) will cause inflammation around TMI requires a separate study.
Although in vitro studies are valuable, the results should be evaluated cautiously. One of the limitations in analyzing the results of this in vitro study was using mouthwashes in a static condition while in a realistic situation; TMI is in contact with mouthwash several times per day. Furthermore, more metal could release in real life because of the fluidity of saliva in the mouth and because oxide layers are removed by tooth brushing.
39
Conclusion
The findings of our study showed that exposure time and mouthwash type influenced the pattern of releasing Al and Ti. However, released ions neither exceed the toxic levels nor the daily dietary intake of metal ions suggesting that OralB, Orthokin, and CHX are safe. However, these types of mouthwash were partially enhancing corrosion.
Financial support and sponsorship
Nil.
Conflicts of interest
The authors of this manuscript declare that they have no conflicts of interest, real or perceived, financial or non-financial in this article.
AzeemMSaleemMMLiaquatAUl HaqAUl HamidWMasoodMFailure rates of mini-implants inserted in the retromolar area539JingZWuYJiangWZhaoLJingDZhangNFactors affecting the clinical success rate of miniscrew implants for orthodontic treatment83541AndrucioliMCMatsumotoMASaraivaMCFeresMFigueiredoLCSorgiCASuccessful and failed mini-implants: Microbiological evaluation and quantification of bacterial endotoxine20170631KnutsonKJBerzinsDWCorrosion of orthodontic temporary anchorage devices5006DanaeiSMSafaviARoeinpeikarSMOshaghMIranpourSOmidkhodaMIon release from orthodontic brackets in 3 mouthwashes: An in vitro study7304AhmedMFEl-SayedAKChenHZhaoRYusufMSZuoQComparison between curcumin and all-trans retinoic acid in the osteogenic differentiation of mouse bone marrow mesenchymal stem cells415466TrolićIMTurcoGContardoLSerdarevićNLOtmačićHCorrosion of nickel-titanium orthodontic archwires in saliva and oral probiotic supplements.Acta Stomatologica Croatica316SzuhanekCGrigoreADetermination of microelements from orthodontic implants by the flame atomic absorption spectroscopy method16002Al-MayoufAAl-SwayihAAl-MobarakNAl-JababACorrosion behavior of a new titanium alloy for dental implant applications in fluoride media3209HuangGYJiangHBChaJYKimKMHwangCJThe effect of fluoride-containing oral rinses on the corrosion resistance of titanium alloy (Ti-6Al-4V)30612KravitzNDKusnotoBRisks and complications of orthodontic miniscrewsS4351ChinMYSandhamAde VriesJvan der MeiHCBusscherHJBiofilm formation on surface characterized micro-implants for skeletal anchorage in orthodontics203240NatarajanMPadmanabhanSChitharanjanANarasimhanMEvaluation of the genotoxic effects of fixed appliances on oral mucosal cells and the relationship to nickel and chromium concentrations: An in vivo study3838GheysarzadehAYazdanparastRSTAT5 reactivation by catechin modulates H2O2-induced apoptosis through miR-182/FOXO1 pathway in SK-N-MC cells64956GheysarzadehAYazdanparastRInhibition of H2O2-induced cell death through FOXO1 modulation by EUK-172 in SK-N-MC cells4752SetcosJCBabaei-MahaniADi SilvioLMjörIAWilsonNHThe safety of nickel containing dental alloys11638KasprzakKSSunderman FWJr, Salnikow K.Nickel carcinogenesis6797CortadaMGinerLCostaSGilFJRodríguezDPlanellJAGalvanic corrosion behavior of titanium implants coupled to dental alloys28793TavakoliLYaminiYEbrahimzadehHNezhadaliAShariatiSNourmohammadianFDevelopment of cloud point extraction for simultaneous extraction and determination of gold and palladium using ICP-OES73743O'ReillyMOral care of the critically ill: A review of the literature and guidelines for practice10110AminiFJafariAAminiPSepasiSMetal ion release from fixed orthodontic appliances—an in vivo study.The European Journal of Orthodontics12630Martín-CameánAMolina-VillalbaIJosAIglesias-LinaresASolanoECameánAMGilFBiomonitorization of chromium, copper, iron, manganese and nickel in scalp hair from orthodontic patients by atomic absorption spectrometry.Environmental toxicology and pharmacology75971MikulewiczMWołowiecPJaneczekMGedrangeTChojnackaKThe release of metal ions from orthodontic appliances animal tests6739MikulewiczMChojnackaKWoźniakBDownarowiczPRelease of metal ions from orthodontic appliances: An in vitro study27280Martín-CameánAJosAPuertoMCallejaAIglesias-LinaresASolanoEIn vivo determination of aluminum, cobalt, chromium, copper, nickel, titanium and vanadium in oral mucosa cells from orthodontic patients with mini-implants by Inductively coupled plasma-mass spectrometry (ICP-MS)1320HouXAmaisRSJonesBTDonatiGLInductively coupled plasma optical emission spectrometry.Encyclopedia of Analytical Chemistry: Applications, Theory and Instrumentation125AbboodiHHAl-DabaghDJAssessment of metal ions released from orthodontic mini-implants in fluoridated mouthwashes15664SerraGMoraisLEliasCNSemenovaIPValievRSalimgareevaGNanostructured severe plastic deformation processed titanium for orthodontic mini-implants4197202CaetanoaPLBahiaaMSe SilvabEdVitralaRWda Silva CamposaMJCorrosion resistance and surface characterization of miniscrews removed from orthodontic patients17MeetingJFWECoFAWorld HealthOrganizationSafety Evaluation of Certain Food Additives and Contaminants: World Health Organization; 2014.Safety Evaluation of Certain Food Additives and Contaminants: World Health Organization; 2014MiloševIKapunBSelihVSThe effect of fluoride ions on the corrosion behaviour of Ti metal, and Ti6-Al-7Nb and Ti-6Al-4V alloys in artificial saliva54355NakagawaMMatsuyaSShiraishiTOhtaMEffect of fluoride concentration and pH on corrosion behavior of titanium for dental use156872KangEHParkSBKimHIKwonYHCorrosion-related changes on Ti-based orthodontic brackets in acetic NaF solutions: Surface morphology, microhardness, and element release55560LópezMFGutiérrezAJiménezJAIn vitro corrosion behaviour of titanium alloys without vanadium135964HwangCJShinJSChaJYMetal release from simulated fixed orthodontic appliances38391ArunAMaheshCChemical and biological evaluation of different commercially available metal orthodontic brackets-An in vitro study370PavlicAPerissinottoFTurcoGContardoLSpaljSDo chlorhexidine and probiotics solutions provoke corrosion of orthodontic mini-implants.An in vitro study?137988MandsaurwalaMKaliaAGuptaGHegdeAMirdehghanNComparative evaluation of ion release from orthodontic mini-implants in 2 mouthwashes: An in vitro study17781KerosuoHMoeGKlevenEIn vitro release of nickel and chromium from different types of simulated orthodontic appliances1116