Stress analysis in implant, abutment, and peripheral bone with different restorative crown and abutment materials: A three‑dimensional finite element analysis study
Abstract
Background: Stresses on prosthetic crown directly influences the survival rate of implants hence
it should be considered while selecting prosthetic material.
The aim of the study is to evaluate stress analysis on implant, abutment and peripheral bone with
change in different abutment and different crown materials by 3D finite element study.
Materials and Methods: A numerical procedure based on finite element method was adopted to
investigate the influence of different prosthetic materials and abutment materials on stress situation.
Eight different three‑dimensional (3D) models of a bone‑level implant system and an abutment
were created by using the standard tessellation language (STL) data of original implant components.
Combinations included of abutment materials i.e., Titanium (Ti), Polyetheretherketone (PEEK),
Polyetherketoneketone (PEKK), Polymer infilterated hybrid ceramic (TZI) along with different
restoration materials Monolithic Zirconia (MZ) and lithium disilicate glass ceramic (IPS e‑max).
In each model, the implants were loaded obliquely (150 N). The stress distribution in the implant,
Abutment and peripheral bone was evaluated through the von Mises stress analysis.
Results: Higher stresses were found on neck of implants irrespective of abutment material and
restorative material. Highest stress was found with PEEK material. The pattern of stress distribution
in implant and peripheral bone was similar in all models.
Conclusion: There is no difference in stresses with the change in restorative material but the
change in abutment material has effect on stresses on implants.
Key Words: Abutment, finite element analysis, implant, stress
Refbacks